EasyManuals Logo

Agilent Technologies 1100 Series Reference Manual

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 81
Synchronization Lost
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable Causes
• CAN cable disconnected.
• Defective CAN cable.
• Defective main board in another module.
Suggested Actions
✔ Ensure all the CAN cables are connected correctly.
✔ Switch off the system. Restart the system, and determine which module or
modules are not recognized by the system.
✔ Ensure all CAN cables are installed correctly.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals