140 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
Correcting Leaks
1 Remove the front cover.
2 Use tissue to dry the leak sensor area and the leak pan.
3 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4 Replace the front cover.
When required If a leakage has occurred in the flow cell area or at the capillary connections
Tools re quired Tissue
Two 1/4 inch wrenches for capillary connections
Parts required None
Figure 56 Observing for Leaks
Leak sensor