EasyManua.ls Logo

Agilent Technologies 1100 Series - TCP;IP Parameter Configuration

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
First Steps with the Fluorescence Detector 2
1100 Series FD Reference Manual 35
2 Set the FLD setpoints according to Figure 7 (on the local control module
this information is split across separate screens).
3 Start the run.
4 The resulting chromatograms are shown in Figure 8:
5 The excitation maxima is around 250 nm.
Figure 7 FLD Parameters
Figure 8 Biphenyl peak with different excitation wavelengths
In this example
additional excitation
wavelengths (B, C,
D) are used. This will
increase the scan
time and may lower
the performance.
Ex=230 nm
Ex=246 nm
Ex=250 nm
Ex=290 nm
Biphenyl peak

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals