Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 109
Method Settings
Tabl e 20 Method Settings
Setting Value in WLEXTEST Value in WLEMTEST
Peak Width >0.2 min (4 s, standard) >0.2 min (4 s, standard)
Fit Spectral Range OFF OFF
PMT Gain 12 12
Flash Lamp ON ON
Spectrum Range EX 377 - 417 nm step 1 nm EM 330 - 370 nm step 1 nm
Store Spectra All w/o signal All w/o signal
EX Wavelength 350 nm, ON 350 nm, OFF
EM Wavelength 397 nm, OFF 397 nm, ON
Multi WL Settings Multi EX Multi EM