108 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
9 You can now select a stored spectra and evaluate/print them.
The maximum should be found at 397 nm
±3nm.
10 Load the method WLEMTEST. The FLD will change into the multi-excitation
mode and scan in the range of the expected maximum of 350 nm
±20 nm.
11 Repeat steps step 6 to step 8.
12 You can now select a stored spectra and evaluate/print them.
The maximum should be found at 350 nm
±3nm.
Figure 39 Evaluating Online Excitation Spectra
NOTE
The spectra on the control module is a real raw data spectrum. Therefore it looks not that
smooth as on the Agilent ChemStation (uses the spline algorithm).