Agilent Technologies 8890 - Automated Testing; Self-Guided Diagnostics

Agilent Technologies 8890
278 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7Diagnostics
Operation Manual 115
Automated testing
The GC performs continuous, automated testing of the following items. If a failure occurs, an
alert appears on the diagnostic tab. See Figure 27. In addition, an entry is made in the
appropriate log.
See below for a list of the continuous monitoring dialogs:
Detector:
• Supply voltage
• ADC (analog to digital converter) references
• FID flameout
• NPD bead open/short
• Igniter open/short
• Collector short
• Electrometer not plugged in
• TCD filament open or short
• TCD switching valve open or short
• FPD flameout
EPC (Electronic Pneumatic Control) - Inlets, detectors, and other modules.
ADC (analog to digital converter) references
Actuator movements
Thermal:
• Sensor open/short
• Missing heater
• Wrong heater
• Heater current:
• Quiescent
• Leakage
Configuration Mismatch
Self-guided diagnostics
The GC offers several useful diagnostic checks to use when troubleshooting issues with the
GC or method. See below for a list of self-guided diagnostic tests available:
Inlet tests:
• Leak & Restriction test
• Gas Supply Pressure test
• Split Vent Restriction test
• Pressure Decay test
• Septum Purge test