7Diagnostics
Operation Manual 115
Automated testing
The GC performs continuous, automated testing of the following items. If a failure occurs, an
alert appears on the diagnostic tab. See Figure 27. In addition, an entry is made in the
appropriate log.
See below for a list of the continuous monitoring dialogs:
Detector:
Supply voltage
ADC (analog to digital converter) references
FID flameout
NPD bead open/short
Igniter open/short
Collector short
Electrometer not plugged in
TCD filament open or short
TCD switching valve open or short
FPD flameout
EPC (Electronic Pneumatic Control) - Inlets, detectors, and other modules.
ADC (analog to digital converter) references
Actuator movements
Thermal:
Sensor open/short
Missing heater
Wrong heater
Heater current:
Quiescent
Leakage
Configuration Mismatch
Self-guided diagnostics
The GC offers several useful diagnostic checks to use when troubleshooting issues with the
GC or method. See below for a list of self-guided diagnostic tests available:
Inlet tests:
Leak & Restriction test
Gas Supply Pressure test
Split Vent Restriction test
Pressure Decay test
Septum Purge test