5Methods
Operation Manual 81
About gas saver
Gas saver reduces carrier flow from the split vent after the sample is on the column. It applies
to the Split/Splitless, Multimode, and PTV inlets (all modes) and to the split and splitless
modes of the Volatiles Interface. It is most useful in split applications.
Column head pressure and flow rate are maintained, while split vent flows decrease. Flows—
except column flow—remain at the reduced level until you press .
The pulsed modes of the split/splitless, multimode, and PTV inlets are similar except for the
pressure pulse starting at and ending at Pulse time.
To use gas saver
1 Select Method > Inlets.
2 Scroll down until the Gas Saver option is visible.
3 Select the checkbox next to Gas Saver to enable it.
4 Input the Setpoint. The setpoint must be at least 15 mL/min greater than the column flow.
0.32 mm id 3.2 2.6 0.75
0.53 mm id 5.3 4.2 1.5
Table 3 Column size and carrier flow rate (continued)
Column type Column size Carrier flow rate, mL/min
Hydrogen Helium Nitrogen
01234 567 8
10
20
30
40
50
-1-2
01234 567 8
10
20
30
40
50
-1-2

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals