17 Chromatographic Checkout
Operation Manual 237
To Prepare for Chromatographic Checkout
Because of the differences in chromatographic performance associated with different
consumables, Agilent strongly recommends using the parts listed here for all checkout tests.
Agilent also recommends installing new consumable parts whenever the quality of the
installed ones is not known. For example, installing a new liner and septum ensures that they
will not contribute any contamination to the results.
When the GC is delivered from the factory, these consumable parts are new and do not need
replacement.
NOTE
For a new GC, check the installed inlet liner. The liner shipped in the inlet may not be the liner
recommended for checkout.
1 Check the indicators/dates on any gas supply traps. Replace/recondition expended traps.
2 Install new consumable parts for the inlet and prepare the correct injector syringe (and
needle, as needed).
Table 40 Recommended parts for checkout by inlet type
Recommended part for checkout Part number
Split splitless inlet
Syringe, 10-µL 5181-1267
O-ring 5188-5365
Septum 5183-4757
Liner 5190-2295
Multimode inlet
Syringe, 10-µL 5181-1267
O-ring 5188-6405
Septum 5183-4757
Liner 5190-2295
Purged packed column inlet
Syringe, 10-µL 5181-1267
O-ring 5080-8898
Septum 5183-4757

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals