4Touchscreen Operation
54 Operation Manual
For details on running methods, see “Running Methods”.
Status/control tray
The status/control tray provides details on the current status of the GC, the current sequence
and method (if connected to an Agilent data system), the time remaining for the current
operation being performed by the GC, run controls, and so on.
The status/control tray is color coded to reflect the run or ready status of the GC:
Green - Ready for a run
Yellow - Not Ready or Shutdown
Blue - Run in Progress
Purple - Preparing Sample
Teal- Sleep mode
Red- Error
Any Early Maintenance Feedback (EMF) flags are also displayed. See “Early Maintenance
Feedback (EMF).
The tray can be expanded by selecting the up arrow on the tray. See Figure 11.
Figure 11. Status/control tray - expanded
The tray can be minimized by selecting the arrow on the tray.
The Stop control immediately terminates the run. If the
GC is in the middle of a run, the data from that run may be
lost.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals