5Methods
Operation Manual 77
3 Navigate to the Home view and select Prep Run . See “Run controls” for more
information.
4 Wait for status Ready to display.
5 Insert the syringe needle through the septum and all the way into the inlet.
6 Simultaneously depress the syringe plunger to inject the sample and select Start .
To run a method to process a single ALS sample
1 Prepare the sample for injection.
2 Load the sample vial into the assigned location in the ALS tray or turret.
3 Load the desired method. (See “Load a Method”.)
4 Navigate to the Home view and select Start to initiate the ALS syringe cleaning, sample
loading, and sample injection method. After the sample is loaded into the syringe, the
sample is automatically injected when the GC reaches the ready state. See “Run controls”
for more information.
To abort a method
1 Select Stop .
2 When you are ready to resume running analyses, load the appropriate sequence or
method. (See “Load a Method”.)