1 Getting Started
28 Operation Manual
GC Status
When the GC is ready to begin a run, the touchscreen shows STATUS: READY FOR INJECTION.
Alternately, when a component of the GC is not ready to begin a run, the touchscreen shows
STATUS: NOT READY and the status indicator on the GC front panel is yellow. Tapping the
Diagnostics tab displays indications of why the GC is not ready.
Alert tones
The GC provides information via beeps.
A series of warning beeps sounds before a shutdown occurs. The GC starts with one beep. The
longer the problem persists, the more the GC beeps. After a short time the component with the
problem shuts down, the GC emits one beep, and a brief message is displayed. For example, a
series of beeps sounds if the inlet gas flow cannot reach setpoint. The message Inlet flow
shutdown is briefly displayed. The flow shuts down after 2 minutes.
WARNING
Before resuming GC operations, investigate and resolve the cause of the hydrogen
shutdown.
One beep sounds when a problem exists, but the problem will not prevent the GC from
executing the run. The GC will emit one beep and display a message. The GC can start the run
and the warning will disappear when a run starts.
Fault messages indicate hardware problems that require user intervention. Depending on the
type of error, the GC emits no beep or a single beep.
Error conditions
If a problem occurs, the GC status bar changes to Not Ready, the GC status indicator turns
yellow, a number appears next to the Diagnostics tab, and the status tray and Diagnostics tab
list the condition that is causing the Not Ready condition. Select Diagnostics to view the issue
and resolve it.
When configured for enhanced GC-MS or GC-HS communications, error messages from the
connected instruments will also be displayed.
Clearing a shutdown condition
When a component is shut down, the GC becomes Not Ready, the status indicator and status
bar turn yellow, and the Diagnostics tab and status tray display messages about the condition
that caused the shutdown.
To clear a shutdown, select Diagnostics > Clear shutdown - ON to turn on all zones, including
the one that was shut down, or select Clear shutdown - OFF to turn on all zones except the one
that was shut down. Note that if you clear the shutdown but do not resolve the problem that
caused it (for example, changing a gas supply cylinder or fixing a leak), the GC will shut down
again.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals