EasyManua.ls Logo

Agilent Technologies 8890 - About Configuration

Agilent Technologies 8890
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Configuration
160 Operation Manual
About Configuration
A devices configuration properties are constant for an instrument hardware setup, unlike
method settings which can change from sample run to sample run. Two example
configuration settings are the gas type flowing through a pneumatic device and the operating
temperature limit of a device. The GC will try to set any configuration properties necessary for
any device that it finds installed. For example, if the GC knows it is connected to an MSD, it will
automatically configure the MSD heated transfer line. The only settings that need to be
configured manually are items that can be changed depending on your analysis, such as
installed columns and carrier gas type, or an item which may be installed in a non-standard
location. For example, an auxiliary EPC module installed in an inlet EPC module position may
require manual configuration.
Ensure all devices have been configured correctly prior to running any samples.

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals