EasyManuals Logo

Agilent Technologies 8890 User Manual

Agilent Technologies 8890
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
10 Settings
154 Operation Manual
To change the miscellaneous settings
The miscellaneous settings include whether to load the last loaded method when the GC is
powered on and whether to run an automated split vent trap test run on a periodic basis.
1 From the System Settings page, select the Miscellaneous page selection button. The
Miscellaneous page appears. See Figure 52.
Figure 52. Miscellaneous page
2 Select the Load Last Loaded Method on Power On check box to load the last loaded
method when the GC is powered on.
3 Select the Run Split Vent Trap Test after 100 run and 30 minutes of sequence inactivity
check box to cause this automated test to run as indicated. (This test takes approximately
two minutes to run.)
4 Select Apply. The GC saves all of the changes made.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals