4Touchscreen Operation
Operation Manual 67
Touchscreen Functionality When the GC Is
Controlled by an Agilent Data System
When an Agilent data system controls the GC, the data system defines the setpoints and runs
the samples. If configured to lock the touchscreen, the data system can prevent the changing
of setpoints.
When an Agilent data system controls the GC, the touchscreen can be used:
• To view run status by selecting the Home view
• To view the method settings by using the Method view
• To display the last and next run times, the run time remaining, and the post-run time
remaining
• To abort a run
Stopping a run using the touchscreen or Browser Interface immediately ends the run. The data
system may retain the data already collected, but no further data is collected for that sample.
Agilent data systems may allow the next run to begin, depending on the data system and its
settings for handling errors.