8 Maintenanceï€
Operation Manual 125
Available Counters
Table 14 lists the most common counters available. The available counters will vary based on
the installed GC options, consumables, and future updates.
Table 14 Common EMF counters
GC
Component
Parts with a counter Type Default value
Detectors
FID Collector Number of injections
Jet Number of injections
Ignitor Number of ignition attempts
TCD Switching solenoid On time
Filament on time On time
ECD Insert liner Number of injections
Time since wipe test On time 6 months
NPD Bead Number of injections
Collector Number of injections
Bead baseline offset pA Value
Bead baseline voltage Voltage value Blos bead: 1.045
Bead current integral pA-sec Value
Bead on time On time Blos bead: 2400 h
FPD+ Ignitor Number of ignition attempts
PMT Number of injections
PMT On time 6 months
Inlets
SSL Gold seal Number of injections 5000
Gold seal Time 90 days
Liner Number of injections 200
Liner Time 30 days
Liner O-ring Number of injections 1000
Liner O-ring Time 60 days
Septum Number of injections 200
Split vent trap Number of injections 10,000
Split vent trap Time 6 months
MMI Liner Number of injections 200
Liner Time 30 days