8 Maintenance
Operation Manual 129
To Enable, Reset, or Change a Limit for an EMF
Counter
When using the GC without a data system, enable or change the limit for a counter as follows:
1 Locate the counter you wish to change. See “Viewing Maintenance Counters”.
2 Select the Component listing for the counter you want to change. See Figure 35.
Figure 35. Settings dialog box
3 To modify a threshold:
a Select the threshold entry. A data entry dialog box appears.
b Enter the desired value. See “Default Thresholds”.
4 To enable or disable a warning, select or deselect Enable for the corresponding counter.
5 Select Apply. The dialog box closes. The entered value is displayed in the corresponding
field.
6 To reset the counter:
a Select Reset Counter. A confirmation dialog box appears.
b Select Yes. The confirmation dialog box closes.
7 Select Apply.