11 Configuration
Operation Manual 183
Another advantage of enhanced communications is the convenience provided at a system
level:
Consolidated EMF tracking
Synchronized instrument clocks (requires Agilent data system)
Synchronized instrument schedules (sleep/wake)
Pass-through display of connected instrument errors to the GC display
See the Installation and First Startup manual for configuration details.
To enable or disable HS communications
To temporarily disable GC-HS communications:
1 Select Settings > Configuration > Headspace.
2 Use the down arrow to scroll until the Enable Communication checkbox is visible.
3 Select or deselect the checkbox next to Enable Communication to enable or disable
communication with the headspace.