17 Chromatographic Checkout
256 Operation Manual
To Check FPD+ Performance (Sample 5188-5953)
To check FPD+ performance, first check the phosphorus performance, then the sulfur
performance.
Preparation
1 Gather the following:
• Evaluation column, HP-5 30 m × 0.32 mm × 0.25 µm (19091J-413)
• FPD+ performance evaluation (checkout) sample (5188-5953), 2.5 mg/L ( 0.5%)
methylparathion in isooctane
• Phosphorus filter
• Sulfur filter and filter spacer
• 4-mL solvent and waste bottles or equivalent for autoinjector.
• 2-mL sample vials or equivalent for sample.
• Chromatographic-grade isooctane for syringe wash solvent.
• Inlet and injector hardware See “To Prepare for Chromatographic Checkout.”
2 Verify the following:
• Chromatographic-grade gases plumbed and configured: helium as carrier gas, nitrogen,
hydrogen, and air.
• Empty waste vials loaded in sample turret.
• 4-mL vial with diffusion cap filled with isooctane and inserted in Solvent A injector
position.
3 Replace consumable parts (liner, septum, traps, syringe, and so forth) as needed for the
checkout. See “To Prepare for Chromatographic Checkout.”
4 Verify that the Lit Offset is set appropriately. Typically, it should be about 2.0 pA for the
checkout method.
5 Install the evaluation column.
6 Set the oven, inlet, and detector to 250 °C and bake out for at least 15 minutes.
7 Be sure to configure the column.
.
Phosphorus performance
1 If it is not already installed, install the phosphorus filter.
2 Create or load a method with the parameter values listed in Table 46.