17 Chromatographic Checkout
Operation Manual 257
Table 46 FPD+ Checkout Conditions (P)
Column and sample
Type HP-5, 30 m × 0.32 mm × 0.25 µm
(19091J-413)
Sample FPD checkout (5188-5953)
Column mode Constant flow
Column flow 6.5 mL/min
Split/splitless inlet
Temperature 180 °C Split/splitless
Mode Splitless
Purge flow 60 mL/min
Purge time 0.75 min
Septum purge 3 mL/min
Multimode inlet
Mode Splitless
Inlet temperature 75 °C
Initial time 0.1 min
Rate 1 720 °C/min
Final temp 1 250 °C
Final time 1 5.0 min
Purge time 1.0 min
Purge flow 60 mL/min
Septum purge 3 mL/min
Purged packed column inlet
Temperature 180 °C
Septum purge 3 mL/min
Cool on-column inlet
Temperature Track oven
Septum purge 15 mL/min