17 Chromatographic Checkoutï€
236 Operation Manual
About Chromatographic Checkout
The tests described in this section provide basic confirmation that the GC and detector can
perform comparably to factory condition. However, as detectors and the other parts of the GC
age, detector performance can change. The results presented here represent typical outputs
for typical operating conditions and are not specifications.
The tests assume the following:
• Use of an automatic liquid sampler. If not available, use a suitable manual syringe instead
of the syringe listed.
• Use of a 10 µL syringe in most cases. However, a 5 µL syringe is an acceptable substitute.
• Use of the septa and other hardware (liners, jets, adapters, and so forth) described. If you
substitute other hardware, performance can vary.