17 Chromatographic Checkoutï€
Operation Manual 255
8 Start the run.
If performing an injection using an autosampler, start the run using the data system or
press .
If performing a manual injection (with or without a data system):
a Select to prepare the inlet for splitless injection.
b When the GC becomes ready, inject 1 µL of the checkout sample and press .
9 The following chromatogram shows typical results for a new detector with new
consumable parts installed. The Aldrin peak will be missing when using the Japanese
sample 5183-0379.