EasyManuals Logo

Agilent Technologies 8890 User Manual

Agilent Technologies 8890
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #254 background imageLoading...
Page #254 background image
17 Chromatographic Checkout
254 Operation Manual
7 If using a data system, prepare the data system to perform one run using the loaded
checkout method. Make sure that the data system will output a chromatogram.
Detector
Temperature 300 °C
Makeup flow (N2) 25 mL/min (constant + makeup)
Baseline output Should be < 1000 display counts. In Agilent
OpenLAB CDS ChemStation Edition
(< 1000 Hz)
Oven
Initial temp 80 °C
Initial time 0 min
Rate 1 15 °C/min
Final temp 180 °C
Final time 10 min
ALS settings (if installed)
Sample washes 2
Sample pumps 6
Sample wash volume 8 (maximum)
Injection volume 1 µL
Syringe size 10 µL
Solvent A pre washes 2
Solvent A post washes 2
Solvent A wash volume 8
Solvent B pre washes 0
Solvent B post washes 0
Solvent B wash volume 0
Injection mode (7693A) Normal
Airgap Volume (7693A) 0.20
Viscosity delay 0
Inject Dispense Speed (7693A) 6000
PreInjection dwell 0
PostInjection dwell 0
Manual injection
Injection volume 1 µL
Data system
Data rate 5 Hz
Table 45 ECD Checkout Conditions (continued)

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals