17 Chromatographic Checkout
Operation Manual 247
7 If using a data system, prepare the data system to perform one run using the loaded
checkout method. Make sure that the data system will output a chromatogram.
8 Start the run.
If performing an injection using an autosampler, start the run using the data system or
press .
If performing a manual injection (with or without a data system):
a Select to prepare the inlet for splitless injection.
b When the GC becomes ready, inject 1 µL of the checkout sample and select on the
touchscreen.
c The following chromatogram shows typical results for a new detector with new
consumable parts installed.
2 4 6 8
20
30
40
50
60
70
Time (min.)
25 uV
C14 C15 C16C14