Lesson 1 Analog Modules
128
Unit Trigger
The XX+ pin in each unit can be the unit trigger pin. This example
is shown in Trigger System in MCA Module (the setup is an
example) on page 126. The unit trigger can trigger its own unit
only. When you use the unit trigger for a unit, the XX+ pin (which
should be connected to a digital channel as the trigger source on
the DUT board) becomes the trigger input use only, throughout the
testflow. Thus, the analog input/output functions on the XX+ pin
cannot be used throughout the testflow.
For using common trigger pins, the full analog input/output
functions are available in each unit.
Pogo Pin Assignments in MCA
The pogo pin assignments of the MCA module are shown in the
table below.
This table also shows the values to be specified in the Mode
column of the Pin Configuration window when you define the MCA
pins (AWG/digitizer/PMU/V-source pins). This number is also
displayed in the Analog Setup window (Hardware Settings page).
The MCA uses two pogo blocks: S2 and S1 pogo blocks. Thus, the
number of pogo pins in the MCA is twice as many as for other
analog modules, where only the S2 pogo blocks are used.
MCA: Pogo Pin Assignments
Pogo Signal GND Mode Pogo Signal GND Mode
Pad Pad Pad Value in Pad Pad Pad Value in
No. (Unit) PinConfig No. (Unit) PinConfig
01 DD- Unit 4 AGND 24 01 HH- Unit 8 AGND 32
02 DD+ (CLK) Unit 4 AGND 20 02 HH+ (CLK) Unit 8 AGND 28
03 CC- Unit 3AGND 23 03 GG- Unit 7AGND 31
04 CC+ (CLK) Unit 3 AGND 19 04 GG+ (CLK) Unit 7 AGND 27
05 BB- Unit 2 AGND 22 05 FF- Unit 6 AGND 30
06 BB+ (CLK) Unit 2 AGND 18 06 FF+ (CLK) Unit 6 AGND 26
07 AA- Unit 1 AGND 21 07 EE- Unit 5 AGND 29
08 AA+ (CLK) Unit 1 AGND 17 08 EE+ (CLK) Unit 5 AGND 25
09 D- Unit 4 AGND 8 09 H- Unit 8 AGND 16
10 D+ Unit 4 AGND 4 10 H+ Unit 8 AGND 12
11 C- Unit 3 AGND 7 11 G- Unit 7 AGND 15
12 C+ Unit 3 AGND 3 12 G+ Unit 7 AGND 11
13 B- Unit 2 AGND 6 13 F- Unit 6 AGND 14
14 B+ Unit 2 AGND 2 14 F+ Unit 6 AGND 10
15 A- Unit 1 AGND 5 15 E- Unit 5 AGND 13
16 A+ Unit 1 AGND 1 16 E+ Unit 5 AGND 9
17 CT1 --- FGND --- 17 CT2 (CLK) --- FGND ---
Orientation
Marker
Signal Pad
Unit 1 to 4: Source units for Source&Measure module, Measure units for Measure module
Unit 5 to 8: Measure units
S1
01
17
S2
S2 Pogo Block S1 Pogo Block
03
05
07
09
11
13
15
4 pins per unit
2 pins for common trigger: CT1/2
Two pogo blocks: S2 and S1
GND Pad
Pogo Blocks
per module

Table of Contents