Sheet #652u – RS232
Version 1.00a User guide ATEQ 6th series Page 3/8
1.1.1. Examples of results frames
Frames for LEAK test:
Pass part frame: <01>:
<01>:30/05/2012 16:52:01
<01>: 487.8 mbar:(OK): 029 Pa
Fail part frame: <01>:
<01>:30/05/2012 16:53:36
<01>: 493.9 mbar:(DT): 114 Pa
Alarm frame: <02>:
<02>:30/05/2012 16:55:24
<02>: 486.4 mbar:(AL): >> F.S. TEST
This frame is of the same type as the print parameters frame except that the different character
strings follow each other and are separated by a punctuation mark which enables the various
boxes to be entered automatically in Microsoft Excel.
This frame is operated by connecting a computer to the instrument's RS 232 link.
Frames for STANDARD SEALED COMPONENTS test:
Pass part frame:
<01>:ABCDEFGHIJKL
<01>:12/04/2013 10:06:10
<01>: 0.997 bar:(PB): 31 Pa
Frames for LARGE LEAK SEALED COMPONENTS test:
Pass part frame:
<01>:ABCDEFGHIJKL
<01>:17/04/2013 15:08:24
<01>:(OK): 0.441 bar
Fail part frame:
<01>:ABCDEFGHIJKL
<01>:12/04/2013 09:48:42
<01>:(AL):SEALED PART VOL TOO SMALL
<01>:ABCDEFGHIJKL
<01>:12/04/2013 09:54:48
<01>:(TD):SEALED PART VOL TOO LARGE
Alarm frame: (learning error) :
<01>:ABCDEFGHIJKL
<01>:12/04/2013 09:21:29
<01>:(AL):SEALED PART LEARN ERROR
Frames for BLOCKAGE test:
Pass part frame P mode:
<03>:P MODE
<03>:02/05/2013 15:17:13
<03>: 0.199 bar:(OK): 0.199 bar
Fail part frame P mode:
<03>:P MODE
<03>:02/05/2013 15:20:07
<03>: 0.110 bar:(TD): 0.110 bar

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATEQ PRIMUS F Series and is the answer not in the manual?

ATEQ PRIMUS F Series Specifications

General IconGeneral
BrandATEQ
ModelPRIMUS F Series
CategoryTest Equipment
LanguageEnglish

Related product manuals