Sheet #622u – Filter
Version 1.04a User guide ATEQ 6th series Page 1/1
FILTER
1. PRINCIPLE
This function enables the slowing down of the sampling speed, by performing an average over
the set measurement time; this makes the displays value more readable during quickly
changing measurements value.
Cette fonction se justifie avec des unités liées au temps (Pa/s, cm
3
/s, l/min, etc.).
2. CONFIGURATION
Activate the function or check if it checked.
Press the key, the cursor slide to the
right hand.
2#4#/2T
(70%6+10
ŹFILTER : No
More functions…
By using the arrows, select
"Yes" and validate with the
key.
2#4#/2T(70%6+10
FILTER : Yes Ż
More functions…
Adjust the Filter (time) parameter.
FILTER time: maximum duration of the
averaging of the measured result.
2#4#/2T
(70%6
(+
FILTER : 2.0 s Ż