Sheet #684u – Error messages
Version 1.04a User guide ATEQ 6th series Page 1/3
ERROR MESSAGES
The ATEQ instrument can display error messages if there are operational problems.
1. COMMUNICATION ERRORS
At starting of the device, the normal message
appears:
Check system.
Please, wait…
Check system.
Please, wait…
If a detection error of one of the component
occurs, the following message appears:
Detection Error
XXXX missing.
Detection error of the sensor board.
Detection Error
SENSOR missing
Detection error of the Inputs / outputs (I/O)
sensor board.
Detection Error
I/O missing
Detection error of the valve codes board.
Detection Error
VALVE C. missing
Turn off and turn on again the device, check if the measurement head is turning on too
(pneumatics valves sounds).
If the issue persists, please contact ATEQ.
If the communication with the sensor board is lost after starting the device, the device will freeze
on the run program number and won't carry on any test cycle.
An error with the sensor board is instantly detected and it needs 30 to 60 seconds to detect an
error with the relay board and valve codes board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATEQ PRIMUS F Series and is the answer not in the manual?

ATEQ PRIMUS F Series Specifications

General IconGeneral
BrandATEQ
ModelPRIMUS F Series
CategoryTest Equipment
LanguageEnglish

Related product manuals