Sheet #631u – Service special cycles
Version 1.04a User guide ATEQ 6th series Page 4/5
1.2.1. 2) Differential sensor
This cycle enables the calibration of the differential sensor. It is important to ensure that the test
pressure is 0.
Press the Start to run the special cycle
.
52'%;%.'
LEAK
Pr 001
LEAK TEST
The device carries out a dump, checks if the
pressure is 0 then carries out a test and
displays the measured pressure.
The verification of the sensor can start.
Stop the special cycle by pressing the RESET
key
.
TEST
0.0
Pa