Sheet #678u – Starting up
Version 1.04a User guide ATEQ 6th series Page 3/4
When the window on the right is displayed, the
instrument is ready to test
Note: the device goes on the last program
carried on.
Pr 001
LEAK TEST
3. SCREEN DÉTAIL
100.0
mbar
OK
0.300
cm3/m
0.000
Pr1
cm3/mn
0.0 s
Green light:
Pass part.
Red light:
Fail Part
Flashing Red light:
Alarm
Green and red lights:
Reworkable Part
Run
pro
g
ram
Measurement
unit
Test result
value
Remaining time
of the current
phase
Vertical lines
test result
Test Result
Pressure
instruction
value
Test Reject
value