Page #74 background imageLoading...
Page #74 background image
Sheet #613u – Sealed components
Version 1.04a User guide ATEQ 6th series Page 8/8
The finalised sealed component parameters
can be reviewed and modified by the user in
the functions menu of the program or by
pressing the
key.
Do the same process for the
FAIL PART
learning cycle (this cycle is optional) with a fail
part inside the chamber.
5'#.'&%1/2
ŹMin Level : 0.010
MEASURE : 0.001
Max Level : 0.020
If the sealed part special cycle is no carried
out before, the alarm message
"Do Learning Cycle" is displayed.
100.0
mbar
ALARM
Do Learning Cycle
SEALED PART
Pr 01 READY