124 BD FACSAria User’s Guide
b Select the voltage field for the parameter you need to adjust.
c To adjust the values you can:
Click the up and down arrows to adjust the voltages in increments
of 1.
Drag the sliders to adjust the voltages.
Press the Ctrl key while clicking the on-screen arrows to increase or
decrease in increments of 10.
Press the Ctrl key while pressing the keyboard arrows to increase or
decrease in increments of 10.
7 Adjust the FSC area scaling until the FSC-A signal matches the FSC-H
signal, if needed.
a Click the Laser tab in the Cytometer window and check at the FSC
Area Scaling value.