EasyManua.ls Logo

BENDIXKing XVUE TOUCH - Figure 6-55 KMG 7010;MD32 Calibration Reading

BENDIXKing XVUE TOUCH
280 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
89000109-002 xVue Touch Installation Manual
Rev 2 Page 6-52
© Honeywell International Inc. Do not copy without express permission of Honeywell.
For Use in Non-Certified Aircraft
(5) Once the specified heading is stable, press the Perform Reading button on the CMT screen.
(6) Verify that the Measured Heading value is within ± 9 degrees:
(a) If the value is within range, press the Confirm Reading button, refer to Figure 6-55.
(b) If the value is out of the range, correct the airplane heading and repeat the steps above,
beginning with Step 4, to re-establish a measured heading value.
(7) Repeat steps 3 through 6 above for each 30 degree increment.
Figure 6-55 KMG 7010/MD32 Calibration Reading

Table of Contents

Related product manuals