CAPINTEC, INC. CAPRAC
®
-t
April 2021 WELL COUNTER INITIALIZATION 6-23
Re-measure
If the measured efficiency is not acceptable, touch the MEASURE SOURCE button to
perform the efficiency measurement again.
Save Efficiency Measurement
If the measured efficiency is acceptable, touch the SAVE EFFICIENCY button. The
screen will appear similar to Figure 6-19 Edit Efficiencies Screen showing the Well
Efficiency(%): field box populated with the measured efficiency value.
Figure 6-19 Edit Efficiencies Screen
Touch the ACCEPT button to accept the displayed value. Figure 6-10 Efficiencies
Setup Screen will re-appear showing the new measured efficiency value.
Touch the CANCEL button to abort any changes. Figure 6-10 Efficiencies Setup
Screen will re-appear showing the previous efficiency value.