CAPINTEC, INC. CAPRAC
®
-t
6-22 WELL COUNTER INITIALIZATION April 2021
Efficiency Results
When the specified counting time is completed, the message “FINISHED” will appear
on the screen below the MEASURE button. The screen will appear similar to Figure
6-17 Spectrum for Efficiency Screen.
The primary photopeak ROI (Energy1(keV)) for the selected nuclide will be
highlighted in red. The total spectrum counts per minute and the total ROI counts per
minute are displayed below the spectrum.
To accept the results of the measurement, touch the ACCEPT button. The efficiency
is calculated and the screen will appear similar to Figure 6-18 Measure Efficiency
Results Screen.
To cancel the results of the measurement, touch the CANCEL button. Figure 6-16
Measure Efficiency Screen will re-appear without saving the measurement.
Figure 6-18 Measure Efficiency Results Screen
Print Results
If a printer is attached to the system, the test results can be printed by touching the
PRINT button.