CAPINTEC, INC. CAPRAC
®
-t
6-26 WELL COUNTER INITIALIZATION April 2021
Element Field
The Element field is required. For the Element name (Element:), the alphanumeric
keypad will appear.
Input the Element name for the nuclide being added (e.g. Cobalt) and touch the
ACCEPT button. Figure 6-21 Add User Nuclide Screen will re-appear with Element:
field box populated with the entered name. The Element name can contain any
combination of 14 alphanumeric characters maximum.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.
Halflife Field
The Halflife field is required. For the Half-life (Halflife:) data, the numeric keypad will
appear along with radio buttons for the time (Year, Day, Hour, Minute, Second).
Input the half-life value using the keypad and touch the appropriate radio button for
the time unit of measure of the nuclide being added and touch the ACCEPT button.
Figure 6-21 Add User Nuclide Screen will re-appear with Halflife: field box populated
with the entered value.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.
Energy1(keV) Field
The Energy1(keV) field is required.
Energy1(keV) is the primary photopeak for the nuclide being added. The peak
efficiency is measured using only the counts recorded in the Region of Interest (ROI)
that surrounds the primary photopeak) and does not use total counts. This allows the
peak finding program to better determine the activity of the nuclide when multiple
peaks or nuclides are present.
For the Energy1(keV) (Energy1(keV):) data, the numeric keypad will appear. Input
the desired keV value by touching the appropriate numbers on the keypad and touch
the ACCEPT button. Figure 6-21 Add User Nuclide Screen will re-appear with
Energy1(keV): field box populated with the entered value.
Note: The minimum Energy keV value is 1.000. The maximum Energy keV value is
2000.000.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.
Energy2(keV) Field
The Energy2(keV) field is not required.