April 2021 WELL COUNTER INITIALIZATION 6-27
Energy2(keV) is the secondary photopeak for the nuclide being added.
For the Energy2(keV) (Energy2(keV):) data, the numeric keypad will appear. Input
the desired keV value by touching the appropriate numbers on the keypad and touch
the ACCEPT button. Figure 6-21 Add User Nuclide Screen will re-appear with
Energy2(keV): field box populated with the entered value.
Note: The minimum Energy keV value is 1.000. The maximum Energy keV value is
2000.000.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.
Energy3(keV) Field
The Energy3(keV) field is not required.
Energy3(keV) is the tertiary photopeak for the nuclide being added.
For the Energy3(keV) (Energy3(keV):) data, the numeric keypad will appear. Input
the desired keV value by touching the appropriate numbers on the keypad and touch
the ACCEPT button. Figure 6-21 Add User Nuclide Screen will re-appear with
Energy3(keV): field box populated with the entered value.
Note: The minimum Energy keV value is 1.000. The maximum Energy keV value is
2000.000.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.
Well Efficiency(%) Field
The Well Efficiency can either be input or measured. A value for the Well Efficiency
must be input or measured before the data can be accepted.
Input Efficiency
To input an efficiency value, touch the Well Efficiency(%): field box. The numeric
keypad will appear. Input the desired efficiency value by touching the appropriate
numbers on the keypad and touch the ACCEPT button. Figure 6-21 Add User
Nuclide Screen will re-appear with selected Well Efficiency(%): field box populated
with the entered value.
Note: The minimum Efficiency value is 0.001. The maximum Efficiency value is
99.999.
To cancel any changes and return to Figure 6-21 Add User Nuclide Screen, touch
the CANCEL button.