CAPINTEC, INC. CAPRAC
®
-t
6-28 WELL COUNTER INITIALIZATION April 2021
Measure Efficiency
To measure the Efficiency, touch the Measure button on the Well Efficiency line.
The procedure for measuring the efficiency is the same as that given in the Measure
Efficiency section on page 6-16.
When all the required data is populated, touch the ACCEPT button to save the data. Figure
6-22 User Nuclides Setup Screen with Nuclide will appear with the added nuclide in the list.
Figure 6-22 User Nuclides Setup Screen with Nuclide
To abort adding a user nuclide, touch the CANCEL button. Figure 6-20 User Nuclides Setup
Screen will re-appear without adding the nuclide.