CAPINTEC, INC. CAPRAC
®
-t
6-48 WELL COUNTER INITIALIZATION April 2021
Count Time
The default Count Time for the selected Wipe Type will be displayed in the Count
Time: field box.
If it is desired to use a different Count Time for this particular location, touch the
Count Time: field box. Figure 6-38 Numeric Keypad Screen will appear.
Input the desired counting time by touching the appropriate numbers on the keypad.
Touch the ACCEPT button to accept the entered value. Figure 6-36 Add New Wipe
Location with Wipe Type Selected Screen will re-appear with the Count Time: field
box for this location populated with the entered value.
Note: The minimum count time that can be input is 30 seconds. The maximum count
time that can be input is 9,999 seconds.
To cancel any changes and return to Figure 6-36 Add New Wipe Location with Wipe
Type Selected Screen, touch the CANCEL button.
Nuclides
Note: Efficiency data must be entered for nuclides to be added. (Refer to the
EFFICIENCIES section on page 6-10)
The default Nuclides for the selected Wipe Type will be displayed in the Nuclides:
field boxes.
The Nuclides section of the screen provides the user with the ability to set nuclides
to look for in Wipes for each specific location.
Nuclides are not required when saving the Wipe Location.
Up to 10 nuclides may be selected to be looked for in a Wipe sample for each
location. The nuclides selected here will only be used for the specific location.
To add or change a nuclide, touch the desired nuclide field box in the Nuclides
section of the screen. The Select Nuclide screen will appear.
The screen displays a listing of nuclides that have an Efficiency value assigned to
them (Refer to the EFFICIENCIES section on page 6-10) – 10 at a time.
The nuclide list is in alphabetical order. The length of the list will vary depending on
which nuclides have an Efficiency value assigned to them (Refer to the
EFFICIENCIES section on page 6-10). The number of pages (or screens) will be
shown in the upper right corner of the screen indicating the length of the list. To scroll
through the list, touch the DOWN ARROW (▼) button. The next group of 10 nuclides
will be displayed and the UP ARROW (▲) button will appear allowing the user to
scroll up in the list.