CAPINTEC, INC. CAPRAC
®
-t
9-24 TESTS April 2021
Duration
The default counting time for each sample is 60 seconds. If it is desired to change the
duration, touch the Duration of each Sample (sec): field box. Figure 9-18 Numeric
Keypad Screen will appear.
Figure 9-18 Numeric Keypad Screen
Input the desired duration (in seconds) by touching the appropriate numbers on the
keypad. Touch the ACCEPT button to accept the entered time. Touch the CANCEL
button to abort any changes.
Counting the background for 60 seconds or longer is recommended.
Note: The minimum duration that can be input is 2 seconds. The maximum duration
that can be input is 9,999 seconds.
Number of Samples
The default number of samples (repetitions) to be counted is 5. If it is desired to
change the number of samples, touch the Number of Samples: field box. Figure
9-18 Numeric Keypad Screen will appear.
Input the desired number of samples by touching the appropriate numbers on the
keypad. Touch the ACCEPT button to accept the entered value. Touch the CANCEL
button to abort any changes.
The recommended number of samples is 10.