CAPINTEC, INC. CAPRAC
®
-t
April 2021 TESTS 9-25
Note: The minimum number of samples that can be input is 5. The maximum
number of samples that can be input is 20.
Measurement
Touch the Measure Cs137 button. Figure 9-19 Chi Square Test Measurement
Screen will appear.
Figure 9-19 Chi Square Test Measurement Screen
To abort the Chi Square Test, touch the CANCEL button. Figure 9-17 Chi Square
Test Screen will re-appear.
To begin the Chi Square Test, place the Cs137 Test Source into the Well Counter
and then touch the MEASURE button.
The system will begin measuring the Cs137 source. As counting progresses, the
screen displays live spectra, measurement time, total cpm, the number of the current
repetition and the total number of repetitions as shown in Figure 9-20 Chi Square
Test Measurement Screen – In Progress.