ALARM SETUP (ALARM LIMITS TABLE)
VF8 DELTA/DELTA XL/KAPPA 5-9
et ENF/i ENF 0 to 7.5 % Off Low 0% High 6%
et HAL/i HAL 0 to 7.5 % Off Low 0% High 6%
et ISO/i ISO 0 to 7.5 % Off Low 0% High 6%
et O
2
10 to 100 % Off Low 10% High 100%
et SEV/i SEV 0 to 9.0 % Off Low 0% High 9%
etCO
2
/etCO
2
* 5 to 95 mmHg (0.7 to 12.6 kPa) Off Low 30 mmHg (4.0 kPa)
High 50 mmHg (6.7 kPa)
FiO
2
18 to 100 % On Low 18% High 100%
GP1/GP2
S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa) Off S: (Adult/Pediatric)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult/Pediatric)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult/Pediatric)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Neonatal)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
HR 20 to 300 beats per minute On Adult:
Low 45 bpm
High 120 bpm
Pediatric:
Low 50 bpm
High 150 bpm
Neonatal:
Low 80 bpm
High 170 bpm
iCO
2
/iCO
2
* 2 to 10 mmHg (0.3 to 1.3 kPa)
(upper high limit only)
Off High: 4 mmHg (0.5 kPa)
ICP/ICP2/
ICP3/ICP4
-25 to 300 mmHg (-3.3 to 40.0 kPa) Off Low 2 mmHg (0.3 kPa)
High 20 mmHg (2.7 kPa)
iO
2
18 to 100 % Off Low 18% High 100%
LA -5 to 300 mmHg (-0.7 to 40.0 kPa) Off Low 0 mmHg (0.0 kPa)
High 20 mmHg (2.7 kPa)
Parameter Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority