5-10 DELTA/DELTA XL/KAPPA VF8
5 A
LARMS
LV S/M/D -5 to 300 mmHg (-0.7 to 40.0 kPa) Off S:
Low 75 mmHg (10.0 kPa)
High 160 mmHg (21.3 kPa)
M:
Low 40 mmHg (5.3 kPa)
High 80 mmHg (10.7 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 25 mmHg (3.3 kPa)
NBP S/M/D Adult: 10 to 250 mmHg
(1.3 to 33.3 kPa)
Pediatric: 10 to 170 mmHg
(1.3 to 22.7 kPa)
Neonatal: 10 to 130 mmHg
(1.3 to 17.3 kPa)
Off S: (Adult)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7 kPa)
S: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Pediatric)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Pediatric)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 80 mmHg (10.7 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 70 mmHg (9.3 kPa)
D: (Neonatal)
Low 25 mmHg (3.3 kPa)
High 60 mmHg (8.0 kPa)
PA S/M/D -5 to 300 mmHg (-0.7 to 40.0 kPa) Off S:
Low 10 mmHg (1.3 kPa)
High 35 mmHg (4.7 kPa)
M:
Low 7 mmHg (0.9 kPa)
High 17 mmHg (2.3 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 13 mmHg (1.7 kPa)
Parameter Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority