ALARM SETUP (ALARM LIMITS TABLE)
VF8 DELTA/DELTA XL/KAPPA 5-11
PLS/PLS* 30 to 300 beats per minute Off Adult:
Low 45 bpm
High 120 bpm
Pediatric:
Low 50 bpm
High 150 bpm
Neonatal:
Low 80 bpm
High 180 bpm
PVC/min Adult and pediatric: 1 to 50 PVC per
minute (upper high limit only)
On High: 10 PVC per minute
RA -5 to 300 mmHg (-0.7 to 40.0 kPa) Off Low 2 mmHg (0.3 kPa)
High 12 mmHg (1.6 kPa)
RESP Adult: 5 to 100 breaths per minute
Pediatric and Neonatal: 5 to 145
breaths per minute
Off Adult:
Low 5 bpm
High 30 bpm
Pediatric/Neonatal:
Low 20 bpm
High 80 bpm
RRc 5 to 145 breaths per minute Adult:
Off
Ped/
Neo:
On
Adult:
Low 5 bpm
High 30 bpm
Pediatric/Neonatal:
Low 20 bpm
High 60 bpm
RRc* 5 to 90 breaths per minute Off Low 5 bpm
High 30 bpm
RV S/M/D -5 to 300 mmHg (-0.7 to 40.0 kPa) Off S:
Low 10 mmHg (1.3 kPa)
High 35 mmHg (4.7 kPa)
M:
Low 7 mmHg (0.9 kPa)
High 17 mmHg (2.3 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 13 mmHg (1.7 kPa)
SpO
2
/SpO
2
2* 20 to 100%
NOTE: High and low SpO
2
(saturation) alarm conditions are
graded “Medium”.
Adult:
Off
Ped/
Neo:
On
Adult/Pediatric:
Low 90% High 100%
Neonatal:
Low 85% High 95%
SpO
2
% 1 to 100% (high limit only) Off Adult/Pediatric:
High 20%
Neonatal:
High 10%
ST Alarms See “ST Alarms Table” on page 10-10.
Parameter Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority