5-12 DELTA/DELTA XL/KAPPA VF8
5 A
LARMS
Modifying Alarm Functions
1. Access the Alarm Limits table (see page 5-6).
2. Scroll to the parameter whose alarm functions you wish to configure and
click.
3. Scroll to the alarm function you wish to modify (the first column, Alarms, is
highlighted when you first click on the parameter).
4. Choose the new setting, and click to confirm your selection.
5. Repeat steps 2 - 4 for each change.
STVM/STVCM 0 to 45 mm (0 to 4.5 mV) See “ST Alarms Table” on page 10-10.
TOF-Cnt 0 to 4 Off Low 0
High 4
TruST Parameters (see: ST TruST Leads)
Temperature
(Ta/b, T2a/b,
T3a/b)
-5.0 to 50.0
o
C (23.0 to 122.0
o
F) Off Low 34.0
o
C (93.2
o
F)
High 39.0
o
C (102.2
o
F)
T/T2/T3 -32.0 to 35.0
o
C (-25.6 to 95.0
o
F) Off Low 0.0
o
C (32.0
o
F)
High 2.0
o
C (35.6
o
F)
tpCO
2
/pCO
2
2* 10 to 150 mmHg (1.3 to 20.0 kPa) On Low 4.0 mmHg (0.5 kPa)
High 6.7 mmHg (0.9 kPa)
tpO
2
/tpO
2
* 10 to 300 mmHg (1.3 to 40.0 kPa) On Low 6.7 mmHg (0.9 kPa)
High 10.7 mmHg (1.4 kPa)
PCCO 0.50 to 25.00 L/min Off Low 5.50 L/min
High 9.50 L/min
PCCI 0.75 to 15.00 L/min Off Low 3.00 L/min
High 8.00 L/min
NOTE:
Turning a parameters alarm function on allows alarms to be triggered for that
parameter. It is not a prerequisite for a parameter to be on display or to be
connected for it to produce an alarm.
Turning a parameters alarm function off prevents the parameter from
triggering an alarm.
Parameter Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority