FIO2 SETUP
VF8 DELTA/DELTA XL/KAPPA 22-5
1 Point Calibration (Room Air)
A 1-point calibration of the sensor to room air (21 % oxygen) should be performed on
a daily basis.
1. Make sure the monitor is turned on and that the NeoMed and the monitor are
set up for FiO
2
monitoring (see page 22-3).
2. Expose the sensor to room air.
3. Open the FIO
2
menu (see page 22-4).
4. Click on
1 Point Cal. The message 21 % Calibration in Progress --
Calibration may take from 1-10 minutes appears.
5. Wait for the message 21 % Calibration Complete to appear. (A message
informs you of a failed calibration. Try calibrating again; if the message
persists, try a new sensor.)
6. Return the sensor to the incubator or oxygen hood.
2 Point Calibration (Cal Gas)
A 2-point calibration of the system, to 100 % dry oxygen and room air (21 % oxygen),
should be performed every week:
1. Make sure the monitor is turned on, that the NeoMed pod is properly
connected, and set up for FiO
2
monitoring.
2. Set up the sensor for O
2
calibration following your hospital’s guidelines.
3. Open the FiO
2
menu (see page 22-4).
4. Click on
2 Point Cal.
5. Supply 100 % O
2
when instructed by the monitor.
6. Click on
Continue. A message appears informing you that the calibration is
in progress and requesting you to wait until calibration is complete before
proceeding with room air calibration.
7. Wait for the system to calibrate. When calibration is finished, the message
100 % Calibration Complete appears.
8. Follow the instructions on page 22-5 to calibrate the system to room air
(1-point calibration).
NOTE: Contact your hospital’s technical personnel for help with 2-point calibration.