MAINTENANCE AND REPAIR
VF8 DELTA/DELTA XL/KAPPA 23-27
Agent? The module has not yet identified or
cannot identify the agent because
of one of the following conditions:
• Unknown agent (for example, not
HAL, DES, ISO, SEV or ENF)
• The agent concentration is too
low
• The vaporizer is leaking
• Traces of disinfectant are present
• (For Scio Four Oxi or Scio Four)
select agent manually.
• Check/replace vaporizer.
etN2O > 82% etN
2
O > 82 % • Check N
2
O concentration in the
fresh gas flow.
• Flush.
MultiGas Sample Line
Occlusion
• The Scio module sample line is
occluded
• The water trap is full, defective, or
not installed.
• Check sample line and replace, if
necessary.
• Check water trap, replace or install,
if necessary.
MultiGas Zero in
Progress
The zeroing of the Scio module is in
progress.
Wait
MultiGas Zero
Accepted
The zeroing was successful. Informational message; no action is
required.
MultiGas Zero Failed • Zero performed using the wrong
gas
• Occlusion or leak present
• HW problem
• Verify that anesthetic gases have
not contaminated the surrounding
atmosphere.
• Check for leaks and occlusions.
• Call the manufacturer.
Check Watertrap/
Sample Line
The water trap is full or the sample
line is blocked.
• Replace the sample line.
• Empty or the replace water trap
(see page 23-25).
MultiGas Too Warm • The fan port is blocked.
• Hardware problem.
• Clear/unblock the port.
• Call the manufacturer.
MultiGas Data Invalid Communication problem • Unplug and re-plug the Scio
module.
• Power-cycle the monitor or undock
and redock monitor.
• Call the manufacturer.
MultiGas Warming Up The Scio module is warming up and
is operating at reduced accuracy
Do not rely on the Scio module
values.
MultiGas
Incompatible
Scio H/W or S/W incompatibility • Check the version numbers.
• Call the manufacturer.
MultiGas
Initialization
The device is initializing. Wait
Message Condition Suggested Action