23-28 DELTA/DELTA XL/KAPPA VF8
23 S
CIO® FOUR MODULES
MultiGas Unplugged The Scio module has become
disconnected.
• Check the connection and
reconnect, if necessary.
• Call the manufacturer.
MultiGas H/W
failure
Loss of communication or hardware
problem.
• Check the connections.
• Unplug and re-plug the Scio
module.
• Power-cycle the monitor or undock
and redock the monitor.
• Call the manufacturer.
Message Condition Suggested Action