B-14 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
Docking Station (mount only)
External Battery
Protection Class: Class 1
Chassis leakage current: <300
µ
A @120 V AC, <500
µ
A @220 V AC
Mode of operation: Continuous
Protection against ingress of
water:
Ordinary (IPX0 per IEC 60529)
Environmental Requirements
Temperature range: Operating: 10 to 40°C (50 to 104 °F)
Storage: –40 to 70°C (–40 to 158 °F)
Relative humidity: Operating: 20 to 95 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure: Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Physical Attributes
Connections: Bedside Docking Station: bedside monitor
Auxiliary Docking Station: R50 Recorder
Size (H x W x D): 75 x 197 x 131 mm (3.0 x 7.8 x 5.1 in)
Physical Attributes
Size (H x W x D):
62 x 182 x 24 mm (2.4 x 7.2 x 0.9 in)
Weight:
635g (1.4 lb)
Electrical Specifications
Battery life:
50 minutes
Charging time:
3.5 hours at 25 °C
Battery type:
Lead-acid
NOTES:
Battery life varies with parameter configuration. The battery life specified above is under the
following load conditions: MultiMed
with SpO
2
sensor, 2 temperature probes, HemoMed pod with
4 IBP transducers and a catheter, NBP taking measurements every 15 minutes, LCD Transport
Brightness at 50%, and no continuous tone being generated.
Battery life may diminish after extended use.
Environmental Requirements
Temperature range:
Operating: 10 to 40 °C (50 to 104 °F)
Storage: –15 to +40 °C (20 to 104 °F)