B-36 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
Respiration
Measurement range --
Beta, Alpha, Theta, Delta Power:
Percentage of total power
Range: 0 - 100%
Resolution:1%
Accuracy:±10%
Measurement range --
burst suppression ratio (BSR):
Percentage of time waveform is isoelectric over last 60s
Range: 0 to100 %
Resolution:1 %
Accuracy:±10 %
Input signal range:
±1.125 mV
Bandwidth: 0.5 to 30 Hz
Baseline recovery time: 5 sec
DC offset:
±200 mV DC
Sampling rate: 8000/sec
EEG scales: 5-500 uV/div
EEG lead labels: None, Left, Right, Front, Back, Right Front, Right Back, Left
Front, Left Back, Fp, Fp1, Fp2, Fz, F3, F4, F7, F8, Cz, C3,
C4, T7, T8, Pz, P3, P4, P7, P8, O, O1, O2
Trending: 24 hours
Each EEG channel for the following parameters:
Median frequency, spectral edge frequency,
total Power, beta, alpha, theta, and delta
powers, BSR
Recordings: R50 recorder:Analog EEG waveforms
Laser printer:Analog EEG waveforms,
Power spectrum
Sensing leads: I or II (user-selectable)
Measurement method: Impedance pneumography
Detection threshold: 0.15 to 4.0 in manual mode (user adjustment)
Adult and Pediatric: 0.20 - 10.5 in auto mode
(automatic adjustment)
Neonatal: 0.20 - 1.5 in auto mode
(automatic adjustment)
Measurement range: 0 - 155 breaths per min
Measurement accuracy: ±1 1/breath/min or 2% of rate (whichever is greater)
Apnea Detection? Adult and pediatric: No
Neonatal: Yes