MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-37
Non-Invasive Blood Pressure (NBP)
Parameter display: Systolic, Diastolic, Mean
Measurement method: Oscillometric technique
Modes of operation: Manual (single measurement), Continuous (5 minutes), or Interval
Interval times: 1, 2, 2.5, 3, 5, 10, 15, 20, 25, 30, 45, 60, 120, and 240 min
Measurement range (heart
rate)
30 to 240 bpm
Pressure measurement
range
(Adult - 270 mmHg):
Systolic NBP: 30 to 250 mmHg
Mean NBP: 20 to 230 mmHg
Diastolic NBP: 10 to 210 mmHg
Pressure measurement
range
(Pediatric - 180 mmHg):
Systolic NBP: 30 to 170 mmHg
Mean NBP: 20 to 150 mmHg
Diastolic NBP: 10 to 130 mmHg
Pressure measurement
range
(Neonatal - 140 mmHg):
Systolic NBP: 30 to 130 mmHg
Mean NBP: 20 to 110 mmHg
Diastolic NBP: 10 to 100 mmHg
Connections: Quick-release hose connector with single airway
Default inflation pressure: Adult (270): 160 mmHg ±10mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Inflation pressure after a
valid measurement
(±10 mmHg):
Adult (270): Previous NBP
SYS
+ 25 mmHg
Pediatric (180): Previous NBP
SYS
+ 25 mmHg
Neonatal (140): Previous NBP
SYS
+ 30 mmHg
Inflation pressure
after an alarm:
Adult (270): 160 mmHg ±10 mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Maximum inflation
pressure:
Adult (270): 265 mmHg ±5 mmHg
Pediatric (180): 180 mmHg ±10 mmHg
Neonatal (140): 142 mmHg ±10 mmHg
Minimum inflation
pressure:
Adult (270): 110 mmHg ±10 mmHg
Pediatric (180): 90 mmHg ±10 mmHg
Neonatal (140): 70 mmHg ±10 mmHg
Maximum measurement
time:
Adult (270): 2 min ±1 sec
Pediatric (180): 2 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Maximum measurement
time including a retry:
Adult (270): 3 min ±1 sec
Pediatric (180): 3 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Software safety cut-off: Adult (270): 273 ±3 mmHg
Pediatric (180): 215 ±3 mmHg
Neonatal (140): 153 ±3 mmHg
Hardware safety cut-off: Adult (270): 300 ±30 mmHg
Pediatric (180): 300 ±30 mmHg
Neonatal (140): 157 ±8 mmHg
Static cuff accuracy: ±3 mmHg
Calibration range: Adult and Pediatric:10 to 260 mmHg ±3 mmHg
Neonatal: 10 to 150 mmHg ±3 mmHg